Simultaneous Estimation of Atorvastatin and Amlodipine by Reverse Phase High Performance Liquid Chromatography
Corresponding Author(s) : V.V. Rajkondawar
Asian Journal of Chemistry,
Vol. 18 No. 4 (2006): Vol 18 Issue 4
Abstract
A simple, accurate, rapid, reproducible HPLC method has been developed for simultaneous estimation of atorvastatin and amlodipine using a mobile phase consisting 0.05 M potassium dihydrogen phosphate and acetonitrile in the ratio of 50 : 50 at a flow rate 1.0 mL/min. A Zorbax XDB 250 x 4.6 mm, C18 column was used as a stationary phase. Quantification was performed using UV detector at 235 nm. The method showed good resolution between two peaks. The result was validated by linearity and recovery studies.
Keywords
Atorvastatin
Amlodipine
HPLC
Rajkondawar, V. (2010). Simultaneous Estimation of Atorvastatin and Amlodipine by Reverse Phase High Performance Liquid Chromatography. Asian Journal of Chemistry, 18(4), 3227–3229. Retrieved from https://asianpubs.org/index.php/ajchem/article/view/20820