Lead(II)-Selective Electrode Based on Dibenzodiaza-15-crown-4
Corresponding Author(s) : Hamid Reza Pouretedal
hrpouretedal@yahoo.com
Asian Journal of Chemistry,
Vol. 16 No. 3 (2004): Vol 16 Issue 3
Abstract
A PVC-based membrane of dibenzodiaza-15-crown-4 reveals a Nernstian potentiometric response for Pb2+ over a wide linear range (5.0 × 10–6–1.0 × 10–2 mol/L) with detection limit of 3.5 × 10–6 mole/L. It has a response time of about 15 s and can be used for at least 3 months without any divergence. The proposed membrane sensor revealed good selectivity for Pb2+ over a wide variety of alkali, alkaline earth and transition metal ions. The electrode was used as an indicator electrode in potentiometric titration of lead.
Keywords
PVC-membrane electrode
Dibenzodiaza-15-crown-4
Lead(II)
(1)
Reza Pouretedal, H.; Hossein Keshavarz, M. Lead(II)-Selective Electrode Based on Dibenzodiaza-15-Crown-4. ajc 2016, 16, 1319-1326.
Download Citation
Endnote/Zotero/Mendeley (RIS)BibTeX