Development of High Pressure Thin Layer Chromatographic Method for Simultaneous Estimation of Ranitidine HCl and Domperidone in Their Combined Dosage Form
Corresponding Author(s) : Manoj S. Charde
Asian Journal of Chemistry,
Vol. 17 No. 4 (2005): Vol 17 Issue 4
Abstract
An accurate and precise high pressure thin layer chromatographic method for simultaneous estimation of ranitidine HCl and domperidone in their combined dosage form has been developed. The study employs a silica gel 60GF254 on aluminium foil and a mobile phase comprising methanol : 1,4-dioxane (4 : 6 v/v). The detection was carried out at 282 nm. The linear detector response for ranitidine HCl was observed between 3.0 to 50 μg/mL while for domperidone 0.2 to 3.5 μg/mL. The recovery study was carried out by standard addition method. The recovery was found to be 100.25 ± 0.859, 100.78 ± 0.592 for ranitidine HCl and 100.29 ± 0.394, 99.54 ± 0.436 for domperidone.
Keywords
Download Citation
Endnote/Zotero/Mendeley (RIS)BibTeX