Quantitative Determination of Anticancer Drug Heptaplatin by HPLC
Corresponding Author(s) : Weiping Liu
liuweiping0917@126.com; xt6207@yahoo.com.cn
Asian Journal of Chemistry,
Vol. 22 No. 10 (2010): Vol 22 Issue 10
Abstract
A method for the direct determination of anticancer drug heptaplatin by HPLC has been described. Phenomenex ODS column (150 mm × 4.6 mm i.d., 5 μm) was used as the stationary phase and a mixture of methanol and water (40:60, v/v) as the mobile phase. The flow rate was 1.0 mL min-1, T = 40 ºC and the detection wavelength was 230 nm. Under these analytical conditions, the linear regression coefficient was found to be 0.9999, the average recovery exceeded 99.9 % with RSD 0.85 %. The method is accurate, reliable and rapid. It can be applied to the quality control of heptaplatin.
Keywords
HPLC
Anticancer drug
Heptaplatin
Analysis
He, X., Zhen, P., Chen, X., Yang, S., Xie, X., & Liu, W. (2010). Quantitative Determination of Anticancer Drug Heptaplatin by HPLC. Asian Journal of Chemistry, 22(10), 8174–8178. Retrieved from https://asianpubs.org/index.php/ajchem/article/view/12091